Physico-Chemical Analysis


Ion Chromatography

Description

Ion chromatography is a precise and sensitive technique used to separate, identify and quantify dissolved ions (anions or cations) in a sample. Separation is based on the elution of ions through an ion-exchange resin column, followed by conductivity detection.

Each ion is characterized by a specific retention time, and the measured conductivity is proportional to its concentration, even at very low levels. Both instruments are equipped with an analytical kit for anions or cations and an autosampler, and are controlled using dedicated software.

Samples are filtered beforehand through a 0.45 µm or 0.2 µm PTFE membrane in the presence of colloids or organic contamination.

IC 850 (Metrohm)

IC 850 (Metrohm)

Applications

  • Cations (lithium, sodium, potassium, magnesium, calcium and strontium) in natural aqueous samples or samples from laboratory experiments.

Dionex Aquion (Thermo Fisher Scientific)

Dionex Aquion (Thermo Fisher Scientific)

Applications

  • Inorganic anions (fluoride, chloride, nitrate, nitrite, bromide, sulfate and phosphate) and organic anions (acetate, formate and propionate) in natural aqueous samples or samples from laboratory experiments.
  • Selenium speciation in porewaters from Callovo-Oxfordian claystone (deep clay formation intended for radioactive waste disposal – Cigéo), in its preserved state and at different degrees of oxidative disturbance.

Organic / Inorganic Carbon Analysis

Multi NC 2100S / Module Liquide (Analytik Jena)

Multi NC 2100S / Module Liquide (Analytik Jena)

Description

The instrument enables the analysis of total carbon (TC), organic carbon (TOC and NPOC) and inorganic carbon (TIC). The technique is based on the conversion of carbon in the sample into CO₂ by thermal combustion at 750 °C using a platinum catalyst. The CO₂ is then transported by a carrier gas (air) to a non-dispersive infrared (NDIR) detector for quantification.

Inorganic carbon is measured after acidification of the sample with 20% H₃PO₄, converting it into CO₂. Non-purgeable organic carbon (NPOC) is measured by first acidifying the sample (pH < 2) with a 2 N HCl solution and then purging it under an air flow.

For dissolved organic carbon measurements, the sample is filtered through a 0.45 µm PTFE membrane before analysis. The instrument is equipped with an automated sample rack (60 positions) and an injection syringe with a variable diameter (0.63 mm and 1 mm), depending on the measurement mode (IC, TC).

Applications

  • Determination of total carbon (TC), organic carbon (COD, TOC) and inorganic carbon (DIC, TIC) from a minimum concentration of 1 mg/L, depending on the sample matrix.
  • Analysis of surface waters, seawater and aqueous liquids from experiments (liquid/liquid extracts, organic compounds from binary and ternary radionuclide/cementitious solution/organic nuclear waste systems).